Model Number 9529
Device Problems Under-Sensing (1661); Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/22/2016
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.The electrogram asystole episode shows unstable signal, noise during asystole, and a regular sinus beat after the pause (rather than a junctional or ectopic beat which would be expected after a real asystole); all of which indicate a dislodged device/false asystole.
 
Event Description
It was reported that implantable cardiac monitor (icm) dislodged and that there was false asystole.The icm remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVEAL XT
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
IPG MFG SWITZERLAND
route du molliau 31
tolochenaz 1131
CH  1131
Manufacturer (Section G)
IPG MFG SWITZERLAND
route du molliau 31
tolochenaz 1131
CH   1131
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5711035
MDR Text Key47191665
Report Number9614453-2016-03499
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K103764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/14/2016
Device Model Number9529
Device Catalogue Number9529
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/07/2016
Initial Date FDA Received06/09/2016
Supplement Dates Manufacturer Received03/07/2016
Supplement Dates FDA Received09/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/22/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age65 YR
-
-