The device and associated accessories were not returned to zoll medical corporation for evaluation.The device history log was returned and evaluated.The device history log shows evidence of a connectivity issue between the electrode pads and the patient, however, this does not necessarily indicate a device malfunction.Throughout the device history log, there is evidence of pacing spikes being delivered with no error messages seen.There are many conditions that can contribute to a loss of ecg via leads.We cannot firmly determine root cause without the evaluation of the product and associated accessories involved.No trend is associated with reports of this type.
|