A patient underwent full prophylactic replacement surgery due to the desire for the latest generator model on (b)(6) 2016.The patient's explanted generator and lead were returned to the manufacturer for product analysis.Abraded openings were noted on the outer and the inner silicone tubing of the explanted lead.No other anomalies, other than those mentioned above, were identified in the returned portion of the lead.Note that since the electrode array portion was not returned for analysis, an evaluation and resulting commentary could not be made on that portion of the lead.During the product analysis, there were no anomalies found with the pulse generator.
|