See mfr.Report #3006443171-2016-00001.The same doctor continued the case with a 2nd handpiece (model number 2000, lot number t150101).The 10cm fibroid was treated numerous times.During one of the ablations, it was noted in the midst of the treatment that a temperature reading was missing from needle #7.The handpiece was retracted after completion of that treatment and inspected both laparoscopically and outside of the patient for potential damage to needle #7.At that time, needle #7 was not extending along with the other 6 needles, indicating a possible breakage in the needle similar to the first handpiece.A third handpiece was pulled from customer stock (model number 2000, lot number t150101) to complete the procedure on the second fibroid.No issues were encountered with the third handpiece.
|