Model Number 2123-0007 |
Device Problems
Fluid/Blood Leak (1250); Split (2537); Torn Material (3024)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/02/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
|
|
Event Description
|
The customer reported that upon opening the door of the pump to troubleshoot ail alarms, a split in the tubing at the silicone segment with leaking was noted in the tubing.There was no patient harm reported.
|
|
Manufacturer Narrative
|
Conclusion code: field left blank- no available code for undetermined or unknown cause.The customer¿s report of leak at silicone segment was confirmed.During visual inspection of the set it was noted that the silicone segment part number had a tear near the upper fitment.The tear was measured to be 0.334mm long.Visual examination under magnification observed a crush mark to the upper blue fitment.No other anomalies were noted.Functional and pressure testing confirmed leaking coming out from the silicone tubing near the upper fitment.No other anomalies were noted on the set.The root cause of the leak was due to a tear in the silicone segment.The cause of the tear is unknown.
|
|
Manufacturer Narrative
|
Additional medwatch information provided.
|
|
Event Description
|
Medwatch: ¿patient was new post-op, new iv fluids hung on alaris pump, no leaks or drips detected.Alaris infusion pump began alarming, it was noted the air was in the line directly after the infusion chamber.When opened the chamber was noted to be wet.Tubing appeared to be split or cracked in soft infusion area.Tubing was saved to be investigated.".
|
|
Search Alerts/Recalls
|