BOSTON SCIENTIFIC - SPENCER RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH Back to Search Results
Model Number M00545000
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/25/2016
Event Type  malfunction  
Manufacturer Narrative
Reported event of brush bent.Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that an rx cytology brush was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2016.According to the complainant, during the procedure, the brush couldn't come out of the sheath and it was bent.The procedure was completed with another rx cytology brush.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Investigation results: detailed analysis of the returned rx cytology brush revealed the unit was consistent with the reported issue of bent brush.The brush wire had been bent in the brush bristle section 5 mm from the distal tip.Visual assessment found the working length kinked.The tuohy-borst connector blue cap had been fully disengaged from the t-fitting threads, and the thumb ring sub-assembly and handle components had been pulled out of the t-fitting.The device handle t-fitting was reassembled for functional evaluation, which found the thumb ring would actuate in both directions without resistance.The brush would fully extend, but could not retract fully into the catheter due to the bent wire at the bristle section.The brush bristles were present and properly formed.The most probable root cause for this event is determined to be operational context since residue on the bristles indicates that the device was used in the procedure and the issue was likely due to procedural/anatomical factors.A review of the device history record (dhr) was performed and no deviations were found.
 
Event Description
It was reported to boston scientific corporation that an rx cytology brush was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2016.According to the complainant, during the procedure, the brush couldn't come out of the sheath and it was bent.The procedure was completed with another rx cytology brush.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX CYTOLOGY BRUSH
Type of Device
ENDOSCOPIC CYTOLOGY BRUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer (Section G)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key5730752
MDR Text Key47591128
Report Number3005099803-2016-01682
Device Sequence Number1
Product Code FDX
Combination Product (y/n)N
Reporter Country CodeGR
PMA/PMN Number
K930348
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date01/31/2018
Device Model NumberM00545000
Device Catalogue Number4500
Device Lot Number18854705
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/25/2016
Initial Date FDA Received06/16/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/02/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/04/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-