Model Number M00545000 |
Device Problem
Bent (1059)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/25/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Reported event of brush bent.Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that an rx cytology brush was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2016.According to the complainant, during the procedure, the brush couldn't come out of the sheath and it was bent.The procedure was completed with another rx cytology brush.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
Investigation results: detailed analysis of the returned rx cytology brush revealed the unit was consistent with the reported issue of bent brush.The brush wire had been bent in the brush bristle section 5 mm from the distal tip.Visual assessment found the working length kinked.The tuohy-borst connector blue cap had been fully disengaged from the t-fitting threads, and the thumb ring sub-assembly and handle components had been pulled out of the t-fitting.The device handle t-fitting was reassembled for functional evaluation, which found the thumb ring would actuate in both directions without resistance.The brush would fully extend, but could not retract fully into the catheter due to the bent wire at the bristle section.The brush bristles were present and properly formed.The most probable root cause for this event is determined to be operational context since residue on the bristles indicates that the device was used in the procedure and the issue was likely due to procedural/anatomical factors.A review of the device history record (dhr) was performed and no deviations were found.
|
|
Event Description
|
It was reported to boston scientific corporation that an rx cytology brush was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2016.According to the complainant, during the procedure, the brush couldn't come out of the sheath and it was bent.The procedure was completed with another rx cytology brush.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|