Follow-up #1: date of submission 07/25/2016.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.Device evaluation: the device has been returned and evaluated by product analysis on 07/18/2016 with the following findings: review of the black box data revealed an unexplained loss of prime event on (b)(6) 2016 at 16:58 followed by ¿power on reset.¿ power to pump was restored (b)(6) 2016 at 16:51.The pump was then manually suspended on (b)(6) 2016 at 16:52 and delivery by the pump was never resumed.On investigation, the pump was powered on and exercised for 24 hours on a 2.0 unit/hour basal rate program.At the conclusion of the exercise, the basal history correctly reflected 2.0 units and the total daily dose showed 48.0 units delivered.The total daily doses added up correctly to reflect the user¿s programmed basal rates.The pump passed delivery accuracy testing without delivery defects.The pump was found to be delivering accurately and within the required range.No delivery interruptions, errors, alarms or warnings occurred during the investigation.Investigation did not duplicate the complaint.(b)(4).
|