MAUDE Adverse Event Report: HOLOGIC, INC THINPREP 5000 PROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC THINPREP 5000 PROCESSOR Back to Search Results
Device Problems Device Operates Differently Than Expected (2913); Data Problem (3196)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  No Answer Provided  
Event Description
A customer in (b)(6) reported one incident where their thinprep 5000 processor with autoloader etched a slide with a mismatched number; without producing an error code.The customer confirmed that the sample was found and reprocessed.The barcode produced was correct and matched the vial id upon processing the sample the second time.Hologic's field service engineer (fse) was dispatched to customer's site.Fse performed tsb-00882.Fse followed up with customer and confirmed instrument is operational since tsb was performed.Although no patients needed to be recalled as a result of this incident this is a reportable event since the instrument did not perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THINPREP 5000 PROCESSOR
Type of Device
THINPREP 5000 PROCESSOR
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer Contact
eva maxwell
250 campus drive
marlborough, MA 01752
MDR Report Key5732114
MDR Text Key47608305
Report Number1222780-2016-00139
Device Sequence Number1
Product Code MKQ
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 06/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/25/2016
Initial Date FDA Received06/17/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-