• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIEBEL FLARSHEIM HUT EXT DR FINAL ASSY-REVERSE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIEBEL FLARSHEIM HUT EXT DR FINAL ASSY-REVERSE Back to Search Results
Model Number HUT EXT DR FINAL ASSY-REVERSE
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/02/2016
Event Type  malfunction  
Manufacturer Narrative
Service spoke with biomed who reported that their site had a power glitch, now the infimed i5 computer powers up into an (b)(6) screen and then will not power up into the application.Field service engineer (fse) went on site and found the computer system hard drive was corrupted after loss of power.Fse utilized `system recovery' in windows to bring the system to a previous stable configuration and computer functioned normally.Fse then verified proper operation according to service checklist (b)(4) and returned the system to the customer fully functional.
 
Event Description
Customer report that during a patient procedure the system failed.Staff had to move the patient to another room to complete the procedure.No reported injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUT EXT DR FINAL ASSY-REVERSE
Type of Device
HUT EXT DR FINAL ASSY-REVERSE
Manufacturer (Section D)
LIEBEL FLARSHEIM
2111 east galbraith road
cincinnati OH 45237
Manufacturer Contact
david benson
2111 east galbraith road
cincinnati, OH 45237
5139485719
MDR Report Key5732400
MDR Text Key48518134
Report Number1518293-2016-00059
Device Sequence Number1
Product Code IXR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Radiologic Technologist
Type of Report Initial
Report Date 06/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHUT EXT DR FINAL ASSY-REVERSE
Device Catalogue Number404007
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/02/2016
Initial Date FDA Received06/17/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-