DEPUY SYNTHES POWER TOOLS COMPACT SPEED REDUCER, 60:1; DRILLS, BURRS, TREPHINES & ACCESSORIES - COMPACT SPEED REDUCER
|
Back to Search Results |
|
Catalog Number CSR60 |
Device Problems
Bent (1059); Partial Blockage (1065); Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was duplicated and confirmed.The assignable root cause could not be determined.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported in the service order from (b)(6) that the compact speed reducer device was not rotating during a procedure.During service and repair evaluation, it was observed that the device gears were damaged, the reduction gear was blocked and the driving shaft was bent.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|