• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS COMPACT SPEED REDUCER, 60:1; DRILLS, BURRS, TREPHINES & ACCESSORIES - COMPACT SPEED REDUCER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS COMPACT SPEED REDUCER, 60:1; DRILLS, BURRS, TREPHINES & ACCESSORIES - COMPACT SPEED REDUCER Back to Search Results
Catalog Number CSR60
Device Problems Bent (1059); Partial Blockage (1065); Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was duplicated and confirmed.The assignable root cause could not be determined.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported in the service order from (b)(6) that the compact speed reducer device was not rotating during a procedure.During service and repair evaluation, it was observed that the device gears were damaged, the reduction gear was blocked and the driving shaft was bent.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPACT SPEED REDUCER, 60:1
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES - COMPACT SPEED REDUCER
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key5736407
MDR Text Key48731479
Report Number1045834-2016-11985
Device Sequence Number1
Product Code HBE
Combination Product (y/n)N
Reporter Country CodeSA
PMA/PMN Number
PK042783
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 06/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCSR60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2016
Initial Date Manufacturer Received 06/02/2016
Initial Date FDA Received06/20/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/26/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-