• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT INFORMATICS CORPORATION STARLIMS PUBLIC HEALTH SOLUTION; CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT INFORMATICS CORPORATION STARLIMS PUBLIC HEALTH SOLUTION; CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE Back to Search Results
Model Number PH 10.06
Device Problem Programming Issue (3014)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/25/2016
Event Type  malfunction  
Manufacturer Narrative
Abbott informatics evaluated and confirmed the issue was reproducible.Upon further investigation, it was determined that this issue is due to the interaction with adobe where the acrrd32.Exxe process is constantly running.Settings were reviewed and the enhanced security setting was disable.Enhanced security settings in adobe was set to enable.This has corrected this event.No additional hang failures have been reported.There are no reports of adverse health consequences due to this issue.
 
Event Description
Starlims public health solution is intended for use by public health laboratory professionals and it is designed to receive and process samples from clinical, culture, environmental, food and other ad-hoc or planned sample submissions.The application provides single and multiple site facilities, the ability to store, retrieve, process and manage the data associated with specimen testing including but not limited to patient demographics, tests ordered, test results, test result interpretation, quality control results, and result codes (flags) that may be associated with the specimen from its entry into the laboratory workflow until the report (including patient report) is generated and released.Sample information may also be retained and used for epidemiological studies or submission to adjunct agencies such as governmental public health organizations.When the software is deployed, configurable settings are programmed based on the customer preferences.For instance, each laboratory enters their specific population based reference ranges that are used to evaluate if a patient's test results are normal, below or above the reference range and require further action such as repeat testing.Each laboratory enters their specific reporting rules for reporting samples through various electronic interfaces.A customer notified abbott informatics that the tuberculosis (tb)-preliminary release by panel takes too long to compile, that it times out.Users have to force reports out multiple times.The reports also go to report delivery queue (rdq) error log.
 
Manufacturer Narrative
The initial report ought to have been 06/21/2016, the date the mdr was submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STARLIMS PUBLIC HEALTH SOLUTION
Type of Device
CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE
Manufacturer (Section D)
ABBOTT INFORMATICS CORPORATION
4000 hollywood blvd
suite 333 south
hollywood FL 33021
Manufacturer (Section G)
ABBOTT INFORMATICS CORPORATION
4000 hollywood blvd
suite 333 south
hollywood FL 33021
Manufacturer Contact
dimitris demirtzoglou
4000 hollywood blvd
suite 333 south
hollywood, FL 33021
9544161651
MDR Report Key5739963
MDR Text Key48822234
Report Number3006130047-2016-00006
Device Sequence Number1
Product Code JQP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Technologist
Type of Report Initial,Followup
Report Date 07/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Medical Technologist
Device Model NumberPH 10.06
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/30/2016
Initial Date FDA Received06/21/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-