The hospital reported that during an endoscopic vein harvesting procedure, ultima opcab system, standard would not lock on to the rail of the retractor.It was set in the lock position but would continue to lift off of the rail.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
|
(b)(4).A lot history record review was completed for lots 25124431, 25123958 and 25123236 the last 3 lots shipped to the account prior to the event date.There was no nonconformance recorded in the lot history.The device was returned to the factory for evaluation.Evidence of blood and clinical use was observed.A visual inspection was performed.The cam and hinge had become misaligned.The device was evaluated for its mechanical function.The device was not able to be assembled onto the reference retractor blade ((b)(4), lot 25092411), therefore it could not initially be evaluate for he locking function.The device was further evaluated with the help of engineering.The device, as received, was noted to have remnants of blood and tissue throughout the device, indicating use of the device within a procedure.The lever of the cam was noted to be out of the proper field of rotation.The device was noted to have worn unlocked position notches on the hinge, indicating that the device was locked / unlocked repeatedly during use (further indicating that the cam lever was in the correct position at some point during the procedure).When trying to load the device onto a maquet blade component (reference retractor blade (b)(4), lot 25092411), it required a great deal of force to slide the device onto and off of the reference rail device.Rather, the device should slide smoothly along the blade rail when in the unlocked position.The device as received was in a perpetually locked state.To analyze the received device further, the cam of the received device was cautiously pushed back into the correct position with the hinge, being careful not to damage the parts.Once the cam lever was back in place, the device was able to be slid on to the reference rail and locked into place with no problems observed.A review of the manufacturing process revealed that each device is evaluated for function prior to release which includes loading the device onto a blade component and locked/unlocked prior to packaging.Each unit must meet all specifications and pass inspection prior to release.Based upon the returned condition of the device, the reported complaint for "would not lock on to the rail of the retractor" could not be confirmed.The event occurred during the use of the device and the procedural operation is unavailable for our review.The device is not intended to be disassembled by the user.The complaint device was correctly positioned and functioned as expected.Therefore, there is no indication that a device defect caused or contributed to the event.
|
The hospital reported that during an endoscopic vein harvesting procedure, ultima opcab system, standard would not lock on to the rail of the retractor.It was set in the lock position but would continue to lift off of the rail.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
|