Device evaluated by mfr.: returned product consisted of a ffr comet pressure wire in one piece.Visual inspection of the device revealed no damage or irregularities.Visual examination of the sensor port showed that the port did have some residual thrombus present.With the residual thrombus inside of the sensor port, this could potentially block the sensor from gathering a signal during the procedure and show an equalizing error like the one the customer witnessed.The pressure wire was connected to the analysis support test bench and all applicable data was correct as designed, there was no difficulty in connecting the wire.There was no evidence of any damage or irregularities contributing to the reported equalizing issues.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|