• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN TOMIOKA CORPORATION AG-920RA; MULTI-GAS UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN TOMIOKA CORPORATION AG-920RA; MULTI-GAS UNIT Back to Search Results
Model Number AG-920RA
Device Problems Device Inoperable (1663); Incorrect Or Inadequate Test Results (2456)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/23/2016
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the multi-gas unit is receiving erratic readings after it is in use for a certain amount of time.He stated he had it calibrated before use and he was using new dry line kits and the auxiliary cable is fully connected.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The customer reported that the multi-gas unit is receiving erratic readings after it is in use for a certain amount of time.He stated he had it calibrated before use and he was using new dry line kits and the auxiliary cable is fully connected.
 
Manufacturer Narrative
The customer's complaint was confirmed.The gas sensing unit was changed and all other functions were tested prior to shipping.Nihon kohden will submit a supplemental report in accordance with 21 cfr part 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AG-920RA
Type of Device
MULTI-GAS UNIT
Manufacturer (Section D)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn; susan shadley
tomioka city, japan 370-2 314
JA  370-2314
Manufacturer (Section G)
NIHON KOHDEN CORPORATION
1-31-4 nishiochiai, shinjuku-k
attn: susan shadley
tokyo, japan 161-8 560
JA   161-8560
Manufacturer Contact
susan shadley
1-31-4 nishiochiai, shinjuku-k
attn: susan shadley
tokyo, japan 161-8-560
JA   161-8560
2687133
MDR Report Key5743892
MDR Text Key48682330
Report Number8030229-2016-00272
Device Sequence Number1
Product Code CCK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/22/2016,05/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAG-920RA
Device Catalogue NumberAG-920RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA06/22/2016
Distributor Facility Aware Date05/23/2016
Device Age58 MO
Event Location Hospital
Date Report to Manufacturer06/22/2016
Initial Date Manufacturer Received 06/22/2016
Initial Date FDA Received06/22/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/21/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/22/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-