MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE Back to Search Results
Model Number MODEL 250
Device Problems Computer Operating System Problem (2898); Device Operates Differently Than Expected (2913)
Patient Problem No Code Available (3191)
Event Date 04/02/2016
Event Type  malfunction  
Event Description
During review of the in-house programming history database, it was observed that a faulted system diagnostic test occurred on (b)(6) 2016.The device was interrogated then programmed.However, the off time of 60 minutes was not corrected back.The magnet mode stimulation was also left on.The treating physician was informed of the software error and the importance reprograming the device to intended settings when necessary.The physician was informed to verify that the patient's magnet mode output current is 0 ma for patients receiving vns therapy for depression.
 
Event Description
Patient's settings were adjusted back to intended settings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5744757
MDR Text Key48835311
Report Number1644487-2016-01419
Device Sequence Number1
Product Code MUZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number521381
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received 05/31/2016
Initial Date FDA Received06/23/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/20/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age76 YR
-
-