• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 BARDEX 2-WAY 5CC ALL-SILICONE FOLEY CATHETER; SILICONE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 BARDEX 2-WAY 5CC ALL-SILICONE FOLEY CATHETER; SILICONE CATHETER Back to Search Results
Catalog Number 165816
Device Problem No Flow (2991)
Patient Problem Urinary Retention (2119)
Event Type  Injury  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the catheter would not drain without manual assistance through burping the bag and line manipulation.The urine was backed up reaching volumes of 550ml.
 
Manufacturer Narrative
Received 1 used foley catheter with the drainage bag attached.Per visual evaluation, no damages or obstructions were observed along the catheter.Per the functional test, a flow test was performed.The unit will be considered acceptable if it drains 250cc of water according to the following flow rate: french size unused used 16 fr.64 sec.Max must drain water flowed properly through the funnel of the catheter and without any interruptions or difficulties.The drain was capable to drain the 250cc in 36 seconds.The catheter was found to be within specification.The complaint was unconfirmed as the problem could not be reproduced.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use states the following: ¿caution: do not aspirate urine through drainage funnel wall.Single use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip syringe.Do not use needle.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Visually inspect the product for any imperfections or surface deterioration prior to use." (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX 2-WAY 5CC ALL-SILICONE FOLEY CATHETER
Type of Device
SILICONE CATHETER
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX  85621
Manufacturer (Section G)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX   85621
Manufacturer Contact
janna parks
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key5757279
MDR Text Key48404990
Report Number1018233-2016-00826
Device Sequence Number1
Product Code EZL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040504
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 06/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Catalogue Number165816
Device Lot NumberNGAP4832
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/06/2016
Initial Date FDA Received06/28/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/13/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/19/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-