TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP, SYSTEM 1
|
Back to Search Results |
|
Model Number 801041 |
Patient Problem
No Information (3190)
|
Event Date 12/11/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).This complaint is related to emdr #1828100-2016-00006.On (b)(6) 2015 the large roller pump reported many "5v supply voltage test failure" events.This will cause an alarm to sound, and a red "x" to appear on the pump icon.The 5v is supplied by the power manager through the network interface card (nic) and associated cables.There is no indication that other pumps were having the same issue.A review of the complaint database lists one related complaint (refer to emdr #1828100-2016-00006) since the date the error had occurred.The reported issue could not be duplicated during laboratory evaluation and the root cause remained unknown.
|
|
Event Description
|
During data log review by the investigator, there was a 5 volt (v) supply voltage test failure on the roller pump.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.The service repair technician (srt) could not duplicate the issue, but replaced the pump pod, cable and display cable as a precaution.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|