• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP, SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP, SYSTEM 1 Back to Search Results
Model Number 801041
Patient Problem No Information (3190)
Event Date 12/11/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This complaint is related to emdr #1828100-2016-00006.On (b)(6) 2015 the large roller pump reported many "5v supply voltage test failure" events.This will cause an alarm to sound, and a red "x" to appear on the pump icon.The 5v is supplied by the power manager through the network interface card (nic) and associated cables.There is no indication that other pumps were having the same issue.A review of the complaint database lists one related complaint (refer to emdr #1828100-2016-00006) since the date the error had occurred.The reported issue could not be duplicated during laboratory evaluation and the root cause remained unknown.
 
Event Description
During data log review by the investigator, there was a 5 volt (v) supply voltage test failure on the roller pump.
 
Manufacturer Narrative
The reported complaint was confirmed.The service repair technician (srt) could not duplicate the issue, but replaced the pump pod, cable and display cable as a precaution.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP, SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5758703
MDR Text Key48619505
Report Number1828100-2016-00469
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number801041
Device Catalogue Number801041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/07/2016
Initial Date FDA Received06/29/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/19/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/29/2004
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number1828100-03/27/12-006-C
Patient Sequence Number1
-
-