It was reported that the physician's office was having intermittent communication issues using the programming system.They had to use another programming system to continue the case.For troubleshooting afterwards, the 9v battery was changed out, the system was not plugged into the wall outlet, and parts were exchanged with known working components with no resolution.No single component could be identified as the source of the issue at the time.The suspect programming system (tablet and wand) was received for product analysis on (b)(6) 2016.Product analysis was completed on the returned wand on 06/22/2016.Other than normal wear related observations, no external visual anomaly was identified with the serial data cable or case.Internal visual inspection of the printed circuit board assembly and associated components revealed no anomaly.Continuity testing of the serial data cable and the battery cable passed.Voltage of the returned battery was 8.81v, above the minimum requirement.The device performed according to functional specifications.Product analysis is underway for the returned tablet.
|
Product analysis for the suspect tablet was completed on 07/08/2016.Included in the analysis are the returned tablet attachments, including the usb to db9 serial adapter cable and an sd card.The tablet was returned without an ac adapter, and the main battery was depleted.Visual inspection identified no anomalies.The tablet was powered using a known good ac adapter with no observed anomalies.The tablet booted to the vns welcome screen with no anomalies.An interrogation was attempted but was unable to be completed.The serial cable's db9 hood was opened and showed that the green data+ wire was no longer soldered on the serial cable pcb.After the wire was soldered onto the pcb, interrogation and diagnostics were performed with no further anomalies observed.The determined cause of the observed wire anomaly was mechanical stress.The tablet's main battery was fully recharged successfully using a known good power supply adapter.Interrogation and diagnostic tests were performed successfully with no observed anomalies.The tablet was powered-on continuously using only the main battery for more than an hour with no anomaly.
|