MAUDE Adverse Event Report: OPTONOL LTD. EX-PRESS GLAUCOMA FILTRATION DEVICE; IMPLANT, EYE VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OPTONOL LTD. EX-PRESS GLAUCOMA FILTRATION DEVICE; IMPLANT, EYE VALVE Back to Search Results
Model Number P-50 PL
Device Problem No Flow (2991)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/01/2016
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: the product was returned for analysis.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Results from the product history record review indicated the product met release criteria.There have been no other complaints reported in the lot number.(b)(4).
 
Event Description
Following a glaucoma implantable filtration device (gfd) surgery a surgeon reported the bleb was flat and the gfd was not flowing.The device was explanted and a trabeculectomy was performed.
 
Manufacturer Narrative
The device was received for investigation.During initial inner illumination test, partial blockage of the lumen was found.After cleaning the device the blockage was removed.Light passed through both sides of the restriction unit.Therefore, there is no indication for a manufacturing related factors that could cause the blockage.During production, 100% final inspection is being performed on the entire batch, including visual inspection and inner illumination.If a blockage would be noticed, the product would have been rejected immediately.Having said that, one may conclude that the blockage was formed after the product left the manufacturing plant.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EX-PRESS GLAUCOMA FILTRATION DEVICE
Type of Device
IMPLANT, EYE VALVE
Manufacturer (Section D)
OPTONOL LTD.
kiryat hatikshoret
communication center
neve-ilan 90850
IS  90850
Manufacturer (Section G)
OPTONOL LTD.
kiryat hatikshoret
communication center
neve-ilan 90850
IS   90850
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key5767987
MDR Text Key49497606
Report Number3003701944-2016-00150
Device Sequence Number1
Product Code KYF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K012852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberP-50 PL
Device Catalogue Number60053
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/08/2016
Initial Date FDA Received07/04/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/13/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age88 YR
-
-