Model Number 300-20
Device Problem Scratched Material (3020)
Patient Problem Pain (1994)
Event Date 01/01/2016
Event Type  malfunction  
Event Description
Patient underwent explant due to pain as reported in mfr.Report # 1644487-2016-01153.An analysis was performed on the returned lead portions.The electrodes portion of the leads was not returned for analysis and therefore a complete evaluation could not be performed on the entire lead product.Abraded openings were found on the outer and inner silicone tubing.These abraded openings and the cut ends that were made during the explanted process most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.What appeared to be white deposits were observed in various locations.Eds (energy dispersion spectroscopy ¿ provides chemical or element identity/composition analysis) was performed on the deposit observed on the marked connector boot and identified the deposit as containing silicon, phosphorus, sodium, magnesium, sulphur and calcium.Refer to attached eds sheet for additional information.With the exception of the inner tubing abraded opening, the condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.The setscrew marks found on the connector pins provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, with no discontinuities identified.Based on the findings, there is evidence of an inner tubing abraded opening.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOL LEAD MODEL 300
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5778892
MDR Text Key49831522
Report Number1644487-2016-01535
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/08/2003
Device Model Number300-20
Device Lot Number2802
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2016
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received 06/17/2016
Initial Date FDA Received07/08/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/18/2001
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age23 YR
-
-