It was reported that the patient was referred for a battery replacement due to the battery being at end of service.During pre-op, high impedance was observed during a diagnostic test.Surgery was postponed due to the patient eating prior to the surgery.X-rays were performed and reviewed by the manufacturer.The x-rays showed that the lead connector pin appeared to not be fully inserted.It was also noted that the lead was not placed per labeling.The strain relief bend was present but it was not secured by a tie down.There was not a sufficient strain relief loop and there were no tie downs present.Additionally, a portion of the lead near the electrodes could be not fully assessed.Therefore a potential lead fracture could not be ruled out in this portion of the lead.The surgery was rescheduled.During the surgery the lead pin was fully inserted into the generator and diagnostic testing was performed.The high impedance did not resolve after the pin re-insertion.The lead and generator were then replaced and the surgeon found a lead fracture near the anchor tether.The explanted lead and generator were disposed of following surgery.Therefore product analysis cannot be completed.
|