Device evaluation: the device has been returned and evaluated by product analysis on 12-jul-2018 with the following findings: during investigation, the last basal delivery was on 2018 (b)(6).The last delivered bolus was a 2.85u normal bolus on (b)(6) 2018.The black box showed an unexplained pump reboot on (b)(4) 2018.The pump was powered back on and deliveries resumed.The total daily dose added up correctly and reflect the users programmed basal rates.The pump passed delivery accuracy testing and was found to be delivering accurately and within range.The pump was exercised for 24hrs with a 2.0u/hr basal rate; at end testing the basal history correctly showed 2.0u and the total daily dose showed 48.0u.There were no errors, alarms or warnings or delivery interruptions occurring during the testing.Unable to duplicate the complaint.This report is made under the requirements of the medical device reporting regulations and does not constitute an admission on the part of animas of any deficiency in the performance of the device.
|