Model Number MODEL 250
Device Problems Device Stops Intermittently (1599); Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/07/2016
Event Type  malfunction  
Event Description
It was reported on (b)(6) 2016 that the physician is reporting intermittent communication issues with his programming wand and/or usb cable.The physician states that he has disconnected the usb cable when storing the vns programmer but he states that over the past week his system has been problematic.The wand often fails and only seems to work if he pushes reset one or two times.He replaced the battery and it seems to work better if he holds the tablet and usb cable very still.The issue resolved when the physician replaced the serial cable with a new one.No patients were impacted by this event.The serial cable does not require return, as the failure mode is understood to be a failure of the serial cable associated with a disconnected wire connection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5780643
MDR Text Key49847535
Report Number1644487-2016-01537
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 06/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0.5
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 06/14/2016
Initial Date FDA Received07/09/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/11/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/23/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-