MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS EXPANDABLE INTRODUCER SHEATH SET; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS EXPANDABLE INTRODUCER SHEATH SET; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 916ES23J
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem Vascular Dissection (3160)
Event Date 06/14/2016
Event Type  Injury  
Manufacturer Narrative
According to the instructions for use (ifu), cardiovascular complications, including perforation or dissection of vessels which may require intervention, are potential adverse events associated with the transfemoral transcatheter aortic valve replacement procedure.According to literature review, and as documented in a technical summary written by edwards lifesciences, vascular complications are a well recognized complication of the transfemoral tavr procedure in this elderly population with multiple co-morbidities.Edwards has reviewed many reports, including screening data records and source documentation of vascular complications and has found that the root cause is typically related to a combination of vessel size, tortuosity and calcifications.Although the incidence is decreasing with smaller sheath/delivery system sizes and physician experience, there will continue to be cases in which vascular complications will occur.The thv physician training manuals instruct on procedural considerations for sheath insertion with regards to proper screening critical to reducing vascular complications.The training manual instructs the operator on proper sheath insertion and withdrawal techniques, including pre-dilating the vessel with the edwards dilators, as needed.It also notes that calcification may reduce lumen diameter and limit or prevent transfemoral passage of the devices.The ifu contraindicates patients with ilio-femoral vessel characteristics that would preclude safe placement of sheaths such as severe obstructive calcification or severe tortuosity.Pre-procedure screening and assessment of the femoral/iliac artery internal diameters will enable the clinician to determine if the sapien xt valve can be delivered transfemorally.Assessment of location and amount of circumferential calcium will aid in determining areas of reduced vessel diameters.The operators are trained to measure minimum vessel diameter taking calcium into account.The physician training manual also lists the minimum recommended vessel size for each size device.Despite the best screening tools, a small percentage of patients will have femoral/iliac vessels that are not amenable to the trans-femoral approach or where increased resistance is encountered during insertion of devices.In many cases, the vessel minimum luminal diameter (mld) may be borderline or below the indicated size.In addition, significant calcification and/or tortuosity, not always appreciable on imaging, could be contributing factors to the event.The minimum required vessel diameter for a 16fr esheath is 6.0 mm.The patient¿s access vessel mld was reported to measure 6.1 mm with moderate calcification and tortuosity.In this case, the cause of the iliac artery dissection and resistance inserting the delivery into the esheath cannot be confirmed.However, it is likely that patient factors (borderline access vessel diameter for a 16fr esheath) in combination with vessel calcification and/or tortuosity not appreciated on imaging, could have contributed to the event.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required at this time.
 
Event Description
As reported by our (b)(6) affiliate, a right iliac artery dissection was found upon removal of the 16fr esheath and a patch angioplasty was performed with a prosthetic vessel.The transfemoral tavr procedure was performed via percutaneous access on the right femoral artery.The 22fr dilator was inserted without difficulty.Slight resistance was felt upon insertion of the 16 fr esheath.When a novaflex+ delivery system was inserted, remarkable resistance was noted.A 23 mm sapien xt valve was uneventfully deployed.The delivery system was withdrawn without difficulty.Upon sheath withdrawal, angiography showed a vascular dissection of the right external iliac artery extending to the access site.Ballooning was performed from contralateral approach.As the ballooning was ineffective, a ¿patch angioplasty¿ was performed with a prosthetic vessel.After the treatment, no abnormality was observed on angiography.Per the report, the minimal luminal diameter (mld) of the access vessel measured 6.1 mm with moderate calcification and moderate tortuosity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS EXPANDABLE INTRODUCER SHEATH SET
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
frances preston, rn
1 edwards way
irvine, CA 92614
9492505190
MDR Report Key5780939
MDR Text Key49125499
Report Number2015691-2016-02186
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P130009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date 06/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/16/2017
Device Model Number916ES23J
Device Lot Number60238237
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/14/2016
Initial Date FDA Received07/09/2016
Date Device Manufactured11/19/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-