Model Number WCD 4000
Device Problem Cut In Material (2454)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/07/2016
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (wires exposed) has been confirmed.Upon investigation the cable connecting the distribution node (dn) to the rear therapy electrodes was pulled from the strain relief, damaging wires in the cable.The root cause for the strained cable was physical abuse.No adverse event resulted from the defective electrode belt.
 
Event Description
A us distributor returned electrode belt sn (b)(4) and reported that wires were exposed on the electrode belt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
zachary nelson
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key5786598
MDR Text Key49998791
Report Number3008642652-2016-04655
Device Sequence Number1
Product Code MVK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 07/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2016
Initial Date Manufacturer Received 06/21/2016
Initial Date FDA Received07/12/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/24/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-