Based on the reported information the customer¿s report of catheter ¿rupture¿ was confirmed to be catheter ¿separation¿.Onyx residue was found within the hub of all three marathon microcatheters.In addition, the returned marathon microcatheters were found to be occluded with onyx.The catheters were also found stretched and separated.The tubing material at the distal separated end of the catheters exhibited with stretching and necking which suggest that the catheters separated when exceeding the tensile strength of the tubing material.The separation likely resulted from tensile failure of the catheters during extraction from the treatment site due to the reported entrapment.Therefore the catheter separation was likely a result of being pulled beyond the 20cm limit noted in the instructions for use (ifu).Per the marathon flow directed micro catheter ifu (instructions for use): ¿note: do not apply more than 20 cm of traction to catheter to minimize risk of catheter separation.However, the definitive cause for the catheter ¿entrapment¿ could not be confirmed.The lot history record review of the marathon reported lot number showed no discrepancies that would have contributed to the reported experience.All products are 100% inspected for damages and irregularities during manufacture.In addition, there is no evidence suggesting that the onyx was defective.Per the onyx ifu: ¿difficult catheter removal or catheter entrapment may be caused by one or more of the following factors: long catheterization time; angio-architecture: very distal arteriovenous malformation fed by afferent, lengthened, small, or tortuous pedicles; vasospasm; reflux; injection time.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|