Catalog Number 221750001 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Information (3190)
|
Event Date 07/01/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.Udi: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Pinnacle liner extractor not retracting (teeth piece).
|
|
Manufacturer Narrative
|
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.(b)(4).
|
|
Manufacturer Narrative
|
The device associated with this report was not returned.Review of the provided photographs confirmed nicks on the teeth.The root cause is attributed to wear out.The date code j1004 indicates the instrument was manufactured in october of 2004 and is almost 12 years old.The investigation could not verify or identify any product contribution to the reported event with the information provided.Based on the investigation, the need for corrective action is not indicated.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|