Model Number MMT-523NAL
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Hyperglycemia (1905)
Event Date 06/23/2016
Event Type  Injury  
Manufacturer Narrative
Unit was received with cracked end cap.Unable to perform basic occlusion test, occlusion test, prime/compromised force sensor system test, excessive no delivery test, or displacement test due to cracked end cap.The drive support disk was inspected and no anomaly was noted.Unit was received with broken reservoir tube lip and battery tube threads.Unit was also received with missing o-ring, cracked case at reservoir tube window corners, corroded battery tube, and minor scratches on lcd window.
 
Event Description
The customer reported via phone call that she experienced high blood glucose levels.Her blood glucose level was 524 mg/dl.The drive support cap was popping out.The customer had injections and took a shot to treat her high blood glucose level.Customer was advised to discontinue use of the device and revert to a backup plan.The customer was advised that the device would be replaced and agreed to return the product for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key5798793
MDR Text Key49743640
Report Number3004209178-2016-56952
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 06/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-523NAL
Device Catalogue NumberMMT-523NAL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/24/2016
Initial Date FDA Received07/15/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age60 YR
Patient Weight79
-
-