Device Problem
No Display/Image (1183)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging a display (blank screen) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may result in an inability to use the product which may lead to long term cessation of delivery.
|
|
Manufacturer Narrative
|
Device evaluation: the device has been returned and evaluated by product analysis on 08/10/2016 with the following findings: the pump powered on to a blank display with functional auditory and vibratory alarms.The pump casing was opened and investigation revealed a defective component on the printed circuit board, resulting in the blank screen.Unrelated to the original complaint, investigation revealed that the battery compartment was cracked.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
|