• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION AG-920RA; MULTI-GAS UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION AG-920RA; MULTI-GAS UNIT Back to Search Results
Model Number AG-920RA
Device Problems Calibration Error (1078); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/17/2016
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer reports the multi gas unit displays a calibration error message.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The biomedical engineer reports the multi gas unit displays a calibration error message.
 
Manufacturer Narrative
Manufacturer narrative: the biomedical engineer reports the multi gas unit displays a calibration error message.The unit has been cleaned and evaluated and the reported "cal error" was duplicated.All malfunctioning parts were replaced.The unit was tested per operator's/service manual and the results were recorded on the maintenance check sheet.The unit completed 8 hours of extended testing and operates to manufacturer's specifications.Though the reported issue was confirmed, the unit was sent back to the customer unrepaired due to the customer being non- reachable to approve the repair costs.Nihon kohden will submit a supplemental report in accordance with 21 cfr part 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AG-920RA
Type of Device
MULTI-GAS UNIT
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, japan 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
2687488
MDR Report Key5801591
MDR Text Key50464572
Report Number8030229-2016-00324
Device Sequence Number1
Product Code CCK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 07/17/2016,06/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAG-920RA
Device Catalogue NumberAG-920RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA07/17/2016
Distributor Facility Aware Date06/17/2016
Device Age157 MO
Event Location Hospital
Date Report to Manufacturer07/17/2016
Initial Date Manufacturer Received 07/17/2016
Initial Date FDA Received07/17/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/18/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-