Analysis of instrument data by siemens healthcare diagnostics indicates that the cause for the falsely depressed ferritin (ferr) result was a hook effect.The customer was inquiring about a low unflagged result.The dimension ferr instructions for use (ifu) addresses the potential for hook effect.It states: "one-step sandwich immunometric assays are susceptible to a high-dose "hook effect," where an excess of antigen prevents simultaneous binding of the capture and detection antibodies to a single analyte molecule.Such samples must be diluted and reassayed prior to reporting the results.The ferr method shows no hook effect up to at least 100,000 ng/ml ferritin." in this instance, the patient sample true results exceeded 100,000 and a discrepant result was unflagged for high absorbance.The reagent is performing within ifu claims.No further evaluation of the device is required.
|