It was originally reported the handheld programming computer (hhd) had a dead battery and was never used so it was returned to the manufacturer.It was later reported the physician would have to hold the connector cable into the hhd otherwise the hhd's screen would freeze.The programming wand, hhd, and associated software were returned to the manufacturer for analysis.Analysis for the programming wand was approved on (b)(6) 2016 and it was found that the programming wand failed to communicate during functional testing.The serial cable, which produced the communication errors, had intermittent conductors at the handle location.A known good bench serial cable was substituted and all communication errors cleared.No visual anomalies were identified.Continuity testing of the battery cable passed.After the serial cable was substitutes, the device met specification requirements and performed according to functions specifications.Device analysis is expected for the hhd and associated software; however, this has not been completed to date.
|