Device Problem
Crack (1135)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging the battery compartment was cracked.There was no indication that the product caused or contributed to an adverse event.This complaint is reportable as the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or cartridge compartment.
|
|
Manufacturer Narrative
|
Follow-up #1: date of submission 09/15/2016 device evaluation: the device has been returned and evaluated by product analysis on 08/22/2016 with the following findings: the battery compartment was observed to be cracked to the left of the bumper pad from the threads traveling down to the case seal.Unrelated to the complaint, the display was dim, faded and pink.A crack was also observed between the case seal and display lens corner.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
|