• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES SALZBURG 4.9MM TI LOCKING BOLT 36MM; PIN, FIXATION, SMOOTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES SALZBURG 4.9MM TI LOCKING BOLT 36MM; PIN, FIXATION, SMOOTH Back to Search Results
Catalog Number 459.360VS
Device Problems Difficult to Insert (1316); Fitting Problem (2183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/04/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Due to the reported intra-operative event, the device was not implanted or explanted.The complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.(b)(6).Investigation could not be completed and no conclusion could be drawn as no device was returned.A review of the device history records has been requested and is currently pending completion.Device used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes (b)(4) reports an event in (b)(6) as follows: it was reported that a patient underwent a surgical procedure on (b)(6) 2016 in order to treat a sustained intertrochanteric femoral fracture.During the procedure, the drill bit reportedly interfered with the nail.The drill bit reportedly went through the cortical bone on the opposite side.As a result, the locking bolt also interfered with the nail and could not be successfully placed through the hole.The surgeon abandoned insertion attempts with the locking bolt and completed the procedure.There was no reported surgical delay or additional medical intervention.This report is 2 of 2 for (b)(4).
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.A device history record review was performed for the complaint device lot.Manufacturing location: (b)(6).Manufacturing date: sep 8, 2015.Expiration date: aug 1, 2025.The (b)(4) piece lot was sterilized by supplier, (b)(6).The review showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.No non-conformances were generated during the production of the complaint device.The complaint device is currently in the evaluation process.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.A manufacturing investigation was performed for the subject device (4.9mm ti locking bolt 36mm, part number 459.360vs, lot number 5935878).The device was received with visual deformation of the inner hexagon.The subject device was re-inspected for all the features pertinent to complaint condition statement, ¿then the surgeon tried to insert the locking bolt; but it interfered to the nail and didn¿t go through the hole." the inspection (visual) of the inner hexagon shows a deformation caused by the surgeon.As previously reported, a device history record review was performed for the complaint device lot.Manufacturing location: (b)(4).Manufacturing date: sep 8, 2015.Expiration date: aug 1, 2025.The (b)(4) piece lot was sterilized by supplier, (b)(4).The review showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.No non-conformances were generated during the production of the complaint device.The complaint condition is unconfirmed from a manufacturing perspective.The root cause is attributed to either the device being worn from normal use and servicing but more likely is attributed to the use of the instrument as described by the surgeon and the resulting damage caused by contact of the locking bolt with the nail and possible slippage of the screwdriver.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.9MM TI LOCKING BOLT 36MM
Type of Device
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
SYNTHES SALZBURG
karolingerstrasse
salzburg 5020
AU  5020
Manufacturer (Section G)
SYNTHES SALZBURG
karolingerstrasse
salzburg 5020
AU   5020
Manufacturer Contact
terry callahan
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5806404
MDR Text Key49947991
Report Number3009417901-2016-10024
Device Sequence Number1
Product Code HTY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K970733
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 07/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number459.360VS
Device Lot Number5935878
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/05/2016
Initial Date FDA Received07/19/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received08/11/2016
08/30/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/08/2015
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-