Device Problems Break (1069); Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Device evaluation: the pump has been returned and evaluated by product analysis on (b)(6) 2016 with the following findings: investigation revealed two cracks in the battery compartment.Unrelated to the complaint, investigation revealed that the audio bolus button was torn and punctured.The audio bolus button responded appropriately during testing.
 
Event Description
On (b)(6) 2016, the reporter contacted animas, alleging a casing/condition (cracked/damaged casing) issue.It was reported that the battery compartment was damaged.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or cartridge compartment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key5806605
MDR Text Key50525110
Report Number2531779-2016-16764
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 06/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Other Device ID Number011084040610005121538457716
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Device Age10 MO
Initial Date Manufacturer Received 06/29/2016
Initial Date FDA Received07/19/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/03/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age65 YR
-
-