Device Problem
Crack (1135)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas alleging a cracked battery compartment.There was no reported adverse event associated with this complaint.This complaint is reportable as the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or cartridge compartment.
|
|
Manufacturer Narrative
|
Follow-up #1: date of submission 08/30/2016.Device evaluation: the device has been returned and evaluated by product analysis on 08/05/2016 with the following findings: the battery compartment was found to be cracked from the threads to the case seal left of the grip pad.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
|