It was reported on (b)(6) 2016 from the physician that the patient was seen that day due to increase in seizures.At the patient's last appointment in (b)(6) the device was set to 2.0/20/250/21/1.1, mag 2.25/60/500, autostim 2.25/60/500, with the threshold at 40% and sensitivity to 1 then interrogated it to confirm programming.The physician stated that she interrogated the patient's device on (b)(6) 2016 and found that all output current settings (normal, magnet, and autostim) had been reset to 0ma and she stated there was a message that said "the device has gone into protective mode." she looked at the magnet history and the last magnet swipe recorded was on (b)(6) 2016, so she thinks this may have been when it happened, and patient has likely been without therapy since then.The physician turned the device back on, but at 1.5ma, 1.625 autostim, 1.75 magnet.She wanted autostim to be higher than normal mode, but didn't want the magnet to be a full 0.5 ma higher.All other settings had not been changed and were kept the same.
|