Device Problems
Break (1069); Crack (1135)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.(b)(6).
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging a casing/condition (cracked/damaged casing) issue.It was reported that the battery compartment was damaged.There is no indication that the product issue caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or cartridge compartment.
|
|
Manufacturer Narrative
|
Follow-up # 1 date of submission 9/01/2016.Device evaluation: the device has been returned and evaluated by product analysis on 8/09/2016 with the following findings:
during visual inspection of the pump, it was observed that the battery compartment had two cracks.There was corrosion observed in the battery compartment.A leak test was performed and failed due to the battery compartment leak.The pump was opened and there was no evidence of moisture found.
|
|
Search Alerts/Recalls
|