• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS - HOUSTON PULSE GEN MODEL 106; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS - HOUSTON PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problems Device Operates Differently Than Expected (2913); Device Sensing Problem (2917)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/24/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the patient was seen in an appointment on (b)(6) 2016.At the beginning of the appointment, the generator had settings on a sensitivity level of 3 and threshold of 40%.Autostimulations were noted in the office visit data.The sensitivity setting was attempted to be lowered to values of 1 and 2, but heartbeat was not detected.After placing the sensitivity setting back to 3, the heartbeat was unable to be verified.The output currents were not programmed off prior to attempting to detect the heart rate.The heartbeat verification was reportedly completed successfully in the implant surgery of the generator.The generator dhr was reviewed and found all specifications were met prior to distribution.The review of operations in the dhr showed the generator was a mechanically routed device.No additional pertinent information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5814718
MDR Text Key51065711
Report Number1644487-2016-01661
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 06/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/27/2018
Device Model Number106
Device Lot Number4744
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 06/24/2016
Initial Date FDA Received07/21/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/25/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age35 YR
-
-