• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problems Break (1069); Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event Description
On (b)(6) 2016, the reporter contacted animas, alleging a casing/condition (damaged cap and case) issue.It was reported that the battery compartment was cracked and the battery cap had stripped threads.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no indication that the product caused or contributed to an adverse event.
 
Manufacturer Narrative
Follow-up #1: date of submission 09/08/2016 device evaluation: the device has been returned and evaluated by product analysis on 08/16/2016 with the following findings: during investigation, the battery compartment was found to be cracked.The battery cap was not returned with the pump and a test cap was used for all steps.The test cap secured and detached properly from the pump.Unrelated to the original complaint, the audio bolus button cover was found to be torn in the center but still functional.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key5815562
MDR Text Key50923303
Report Number2531779-2016-17164
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 07/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Other Device ID Number011084040610005121436674016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Device Age15 MO
Initial Date Manufacturer Received 07/03/2016
Initial Date FDA Received07/22/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/08/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age59 YR
-
-