Device Problems
Break (1069); Crack (1135)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging a casing/condition (damaged cap and case) issue.It was reported that the battery compartment was cracked and the battery cap had stripped threads.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no indication that the product caused or contributed to an adverse event.
|
|
Manufacturer Narrative
|
Follow-up #1: date of submission 09/08/2016 device evaluation: the device has been returned and evaluated by product analysis on 08/16/2016 with the following findings: during investigation, the battery compartment was found to be cracked.The battery cap was not returned with the pump and a test cap was used for all steps.The test cap secured and detached properly from the pump.Unrelated to the original complaint, the audio bolus button cover was found to be torn in the center but still functional.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.(b)(4).
|
|