Device Problems
Break (1069); Crack (1135)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.(b)(6).
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging a casing/condition (cracked/damaged casing) issue.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or cartridge compartment.There was no indication that the product caused or contributed to an adverse event.
|
|
Manufacturer Narrative
|
Follow-up #1: date of submission 09/21/2016.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.Device evaluation: the device has been returned and evaluated by product analysis on 08/27/2016 with the following findings: on investigation, moisture was observed in the battery compartment.The battery compartment was cracked from the threads to case seal left of grip pad.Leak testing failed due to a battery compartment leak.There was damage, defect or contamination of the internal pump components.Unrelated to the original complaint, the display was noted to be dim and discolored.
|
|