ABBOTT INFORMATICS CORPORATION STARLIMS PUBLIC HEALTH SOLUTION; CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE Back to Search Results
Model Number PH 10.02
Device Problem Incorrect Software Programming Calculations (1495)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/03/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(6) test configuration was reviewed by customer support and an issue has been found on the configuration of the test.The customer used the incorrect schema.It was misconfigured by the end user.There are no reports of adverse health consequences due to this issue.
 
Event Description
Starlims public health solution is intended for use by public health laboratory professionals and it is designed to receive and process samples from clinical, culture, environmental, food and other ad-hoc or planned sample submissions.The application provides single and multiple site facilities the ability to store, retrieve, process and manage the data associated with specimen testing including but not limited to patient demographics, tests ordered, test results, test result interpretation, quality control results, and result codes (flags) that may be associated with the specimen from its entry into the laboratory workflow until the report (including patient report) is generated and released.Sample information may also be retained and used for epidemiological studies or submission to adjunct agencies such as governmental public health organizations.When the software is deployed, configurable settings are programmed based on the customer preferences.For instance, each laboratory enters their specific population based reference ranges that are used to evaluate if a patient's test results are normal, below or above the reference range and require further action such as repeat testing.Each laboratory enters their specific reporting rules for reporting samples through various electronic interfaces.A customer notified abbott informatics that an (b)(6) result was reported out as reactive for (b)(6) when 2 out of the three tests were (b)(6).The specimen should have resulted out as (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STARLIMS PUBLIC HEALTH SOLUTION
Type of Device
CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE
Manufacturer (Section D)
ABBOTT INFORMATICS CORPORATION
4000 hollywood blvd
suite 333 south
hollywood FL 33021
Manufacturer (Section G)
ABBOTT INFORMATICS CORPORATION
4000 hollywood blvd
suite 333 south
hollywood FL 33021
Manufacturer Contact
dimitris demirtzoglou
4000 hollywood blvd
suite 333 south
hollywood, FL 33021
9544161651
MDR Report Key5817196
MDR Text Key51141533
Report Number3006130047-2016-00008
Device Sequence Number1
Product Code JQP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 07/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberPH 10.02
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/03/2015
Initial Date FDA Received07/22/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-