Catalog Number 03.501.080
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.Additional narrative: patient information is not available for reporting.(b)(6).Device is an instrument and is not implanted/explanted.A device history record review was performed for the subject device lot.The review showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.No non-conformances were generated during the production of the subject device.A service history record review was performed for the subject device.The repair technician reported the lever was sticky and binding.Binding is the reason for repair.The cause of the issue is unknown.No parts were replaced.The device was repaired and will be returned to the customer upon completion.Additional release to warehouse date: january 10, 2014.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was initially reported the application instrument for sternal zipfix does not work.It is unknown if the device was involved in a surgery.It is unknown who discovered the malfunction or how was it discovered.No additional information.Based on investigation received on july 11, 2016, the lever of the sternal zipfix was sticky and binding.The complaint was reassessed for reportability.This is report number 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.No service history review can be performed as device is a lot/batch controlled item.The manufacture date of this item is january 10, 2014.The source of the manufacture date is the release to warehouse date.The service history review is unconfirmed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APPLICATION INSTRUMENT FOR STERNAL ZIPFIX
Type of Device
CERCLAGE FIXATION
Manufacturer (Section D)
SYNTHES HAGENDORF
im bifang 6
hagendorf CH461 4
SZ  CH4614
Manufacturer (Section G)
SYNTHES HAGENDORF
im bifang 6
hagendorf CH461 4
SZ   CH4614
Manufacturer Contact
terry callahan
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5819909
MDR Text Key50355656
Report Number3003875359-2016-10427
Device Sequence Number1
Product Code JDQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 06/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.501.080
Device Lot Number8735273
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/11/2016
Initial Date FDA Received07/25/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/25/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-