• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD ADULT NASAL INTERFACE; CAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD ADULT NASAL INTERFACE; CAT Back to Search Results
Model Number OPT944
Device Problem Device Issue (2379)
Patient Problem Low Oxygen Saturation (2477)
Event Date 06/29/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The opt944 interface is used to deliver humidified oxygen to patients and consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Method: the complaint cannulae were not returned to fisher & paykel healthcare for evaluation.Our investigation is based on a photograph of one of the complaint devices provided by the customer.Results: visual inspection of the photograph revealed that the tubing appeared to have been pulled out of the swivel connector.Conclusion: the observed damage is most likely the result of pulling on the cannula tubing with undue force.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discolouration and stretching or deformation.Any product that fails the visual inspection is disposed of.The setup instructions in the user instructions which accompany the opt944 nasal cannula include the following steps: ensure head strap clip is attached, to prevent cannula from being pulled out of the nares; cannula can become unattached if not used with the head strap clip; attach tubing clip to clothing/bedding to prevent cannula from pulling off face.The user instructions also contain the following warnings/cautions: do not crush or stretch tube, to prevent loss of therapy; failure to use the set-up described above can compromise performance and affect patient safety.Our optiflow production team has been notified of this failure.They have carried out an internal investigation of the manufacturing process and production staff have received additional training to ensure they are following the correct assembly procedure.
 
Event Description
A hospital in (b)(6) reported that an opt944 nasal cannula tube disconnected at the swivel and the patient desaturated.A second cannula was found with swivel detached before patient use.No patient consequence was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT NASAL INTERFACE
Type of Device
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
173 technology drive
suite 100
irvine, CA 92618
9194534000
MDR Report Key5821260
MDR Text Key50389491
Report Number9611451-2016-00368
Device Sequence Number1
Product Code CAT
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 06/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT944
Device Catalogue NumberOPT944
Device Lot Number2100020913
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/30/2016
Initial Date FDA Received07/26/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/25/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-