Device evaluated by manufacturer: the device was returned to edwards for evaluation and the complaint condition was confirmed.As received, the cannula tip was found broken off from the cannula at the suture bump.A cross surface of the broken areas appeared uneven and rough.Edges of broken pieces matched up.No other visible damage, contamination, or other abnormalities were observed from returned device.Manufacturing records were unable to be reviewed as not lot number was provided.A supplier manufacturing defect was not identified.Based on the information received and product evaluation finding, a definitive root cause could not be confirmed.The fmea adequately addressed potential causes of failure and the associated hazards.The complaint trend was assessed and it was found to be in control.No further corrective or preventative actions are required at this time.Trends will continue to be monitored through the use of edwards quality systems and if action is required, appropriate investigation will be performed.
|