• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE ROTALINK¿ PLUS; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE ROTALINK¿ PLUS; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number H749236310020
Device Problem Hole In Material (1293)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/28/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: the device was not returned for analysis.The batch number is unknown and the manufacturing records for the complaint device could not be reviewed.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors. (b)(4).
 
Event Description
It was reported that shaft leak occurred.The target lesion was located at the mid left anterior descending artery.A 1.25mm rotalink plus was selected for use.During the procedure, it was noted that there was a small puncture on the shaft and was leaking rotaglide.The operating team continued with the procedure and covered the puncture with there fingers to prevent further leaking.The procedure was completed with this device.No patient complications were reported and the patients' status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTALINK¿ PLUS
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5823368
MDR Text Key50516770
Report Number2134265-2016-06621
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH749236310020
Device Catalogue Number23631-002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/01/2016
Initial Date FDA Received07/26/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-