• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO FOOTED ATTACHMENT 16MM; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO FOOTED ATTACHMENT 16MM; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Catalog Number 5407FA2000
Device Problems Bent (1059); Flaked (1246)
Patient Problem No Patient Involvement (2645)
Event Date 03/02/2016
Event Type  malfunction  
Manufacturer Narrative
This mdr report is part of the (b)(6) program, (b)(4).One device was received for evaluation.The reported event was confirmed that the device had paint chipping and a bent foot.Paint chipping is known to be caused by non-recommended cleaning practices and/or improper use/handling.A bent foot is known to be caused by excessive side load, which is warned against in the instructions for use.This device is not repairable and was not returned to the user facility.There were no remedial actions taken.This device is not labeled for single-use.
 
Event Description
This report summarizes 1 malfunction event, in which the device was found to have a bent foot during a procedure and was found to have chipped paint when received by stryker.There was patient involvement in regards to the reported bent foot; no patient impact.There was no patient involvement in regards to the disassembly; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOTED ATTACHMENT 16MM
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key5830495
MDR Text Key50720874
Report Number0001811755-2016-01347
Device Sequence Number1
Product Code HBE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number5407FA2000
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/02/2016
Initial Date FDA Received07/28/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse