STRYKER INSTRUMENTS-KALAMAZOO FOOTED ATTACHMENT 16MM; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
|
Back to Search Results |
|
Catalog Number 5407FA2000 |
Device Problems
Bent (1059); Flaked (1246)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/02/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
This mdr report is part of the (b)(6) program, (b)(4).One device was received for evaluation.The reported event was confirmed that the device had paint chipping and a bent foot.Paint chipping is known to be caused by non-recommended cleaning practices and/or improper use/handling.A bent foot is known to be caused by excessive side load, which is warned against in the instructions for use.This device is not repairable and was not returned to the user facility.There were no remedial actions taken.This device is not labeled for single-use.
|
|
Event Description
|
This report summarizes 1 malfunction event, in which the device was found to have a bent foot during a procedure and was found to have chipped paint when received by stryker.There was patient involvement in regards to the reported bent foot; no patient impact.There was no patient involvement in regards to the disassembly; no patient impact.
|
|