STRYKER INSTRUMENTS-KALAMAZOO WIRE COLLET; ARTHROSCOPE Back to Search Results
Catalog Number 4100062000
Device Problem Metal Shedding Debris (1804)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/02/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Approx 4 device was received for evaluation; 4 events were confirmed during testing.Approx 2 devices were found to be affected by corrosion.Approx 1 device was found to be affected by scoring on the driveshaft.Approx 1 device were found to be affected by broken internal component.Approx 1 device was not available to stryker for evaluation.  this device is not repairable and was not returned to the user facility.There were no remedial actions taken.This device is not labeled for single-use.
 
Event Description
This report summarizes 5 malfunction events in which the device produced metal shavings.Approx 1 event had patient involvement; no patient impact.Approx 4 events had no patient involvement; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WIRE COLLET
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key5830960
MDR Text Key51453627
Report Number0001811755-2016-01523
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported5
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number4100062000
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/02/2016
Initial Date FDA Received07/28/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-