• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Bent (1059); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/07/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a physician's programming tablet had bent usb prongs that were preventing communication with the programming wand.The physician had a backup system so no patient therapy was affected.The physician was also provided with a replacement tablet.The programming tablet has been returned to the manufacturer, but product analysis has not been completed to date.The device history record for the tablet was reviewed and showed that there were no non-conforno further relevant information has been received to date.
 
Event Description
Product analysis was completed on the returned tablet and confirmed that there was damage on the tablet usb port and this caused the tablet to be unable to establish communication.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5838112
MDR Text Key51958659
Report Number1644487-2016-01712
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 07/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0.5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 07/07/2016
Initial Date FDA Received08/01/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/11/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/20/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-