STAAR SURGICAL COMPANY MICROSTAAR® INJECTOR; INTRAOCULAR LENS GUIDE Back to Search Results
Model Number MSI-TR
Device Problem Torn Material (3024)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/06/2016
Event Type  malfunction  
Manufacturer Narrative
Expiration date - unk.Date of implant: n/a.Date of explant: n/a.Results: (operational problem): visual inspection of the returned product found the msi-tr injector had no visible damage and there was traces of clear surgical residue.The lens had pieces torn off and missing.The lens was returned dry.(b)(4).
 
Event Description
The reporter stated the surgeon inserted the aa4203tl silicone single piece lens +23.0 se/ 2.0 diopter and noticed the trailing haptic was damaged.The lens was removed and no patient injury was reported.The reporter indicated the injector may be the cause of the lens damage.
 
Manufacturer Narrative
Based on the complaint history and work order search, and product evaluation a specific root cause of the event could not be determined.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROSTAAR® INJECTOR
Type of Device
INTRAOCULAR LENS GUIDE
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
michelle andres
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key5838254
MDR Text Key50903404
Report Number2023826-2016-01002
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K954600
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberMSI-TR
Device Catalogue NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/06/2016
Initial Date FDA Received08/01/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/16/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age64 YR
-
-